6. RESET OF THE INSTRUMENT AND DEFAULT PARAMETERS
BEFORE OF A RESET PROCEDURE DOWNLOAD THE STORED TESTS ON A PC
6.1. RESET PROCEDURE
1. Press the DISP, CLR, RCL keys and then ON/OFF key.
2.
es
The screen alongside is
displayed for 5 seconds, after
which the instrument emits a beep
and then displays the screen
relative to the function selected
with the rotary switch.
ATTENTION: the reset procedure will cancel all the stored tests and it will set the
default parameters in the instrument.
6.2. DEFAULT PARAMETERS
After a RESET the default parameters set on the instrument are:
Function Parameter RESET default parameter
Mode AUTO
Calibration Offset
0
LOW
Mode R+/R- TIMER
TIMER is set at 1s
Mode MAN
Test voltage 500V
R
ISO
Mode TIMER TIMER is set at 60s
EARTH ρ
Parameter DIST DIST = 1
Parameter PLA P = 1
Memory
Memory state 0